PDF(5213 KB)
A 24-bit incremental sigma-delta ADC with a 3rd-order, 4-bit single-loop feedforward structure for bioimpedance measurement
SUN Aoran, CHEN Changhuan, CHEN Yang, SUN Quan, ZHANG Jie, WANG Xiaofei, ZHANG Hong
Integrated Circuits and Embedded Systems ›› 2025, Vol. 25 ›› Issue (2) : 1-10.
PDF(5213 KB)
PDF(5213 KB)
A 24-bit incremental sigma-delta ADC with a 3rd-order, 4-bit single-loop feedforward structure for bioimpedance measurement
Bioimpedance measurement technology serves as a vital tool for medical diagnosis and health monitoring in various emerging medical instruments, in which the high-precision readout circuit is a crucial module for achieving accurate diagnoses. Traditional readout circuits in bioimpedance measurement systems often utilize SAR ADCs, which can hardly achieve an accuracy exceeding 12 bits to meet the increasingly high-precision demands for bioimpedance readout without calibration. To deal with this issue, this paper presents a 24-bit incremental Σ-Δ ADC designed for bioimpedance measurement which adopts a 3rd-order, 4-bit, single-loop feedforward structure to reduce quantization noise while accelerate the conversion speed. The design incorporates data weighted averaging (DWA) technique to mitigate the nonlinearity caused by multibit feedback DAC’s mismatch, significantly enhancing the ADC’s accuracy. In addition, system-level chopping is employed to eliminate systematic offset in the ADC frontend. A configurable digital filter is designed to achieve flexible tradeoff between measurement time and by configuring different oversampling ratios (OSR). The ADC is fabricated using a 180 nm process, with measurement results under an OSR of 128 showing that the ADC achieves an ENOB of 17.80, a SNR of 108.89 dB, a THD of -113.29 dB, an equivalent input RMS noise of 2.95 μVRMS. The power consumption is 930 μW for the whole ADC chip.
bioimpedance measurement / high-order incremental Σ-Δ ADC / multi-bit quantization / dynamic weight averaging algorithm / system-level chopping
| [1] |
杨文卉. 胸腔供血生物阻抗检测与分析方法研究[D]. 沈阳: 沈阳工业大学, 2022.
|
| [2] |
祁朋祥. 基于生物电阻抗的人体成分测试与研究[D]. 合肥: 中国科学技术大学, 2009
|
| [3] |
|
| [4] |
A low-power analog signal processing IC is presented for the low-power heart rhythm analysis. The ASIC features 3 identical, but independent intra-ECG readout channels each equipping an analog QRS feature extractor for low-power consumption and fast diagnosis of the fatal case. A 16-level digitized sine-wave synthesizer together with a synchronous readout circuit can measure bio-impedance in the range of 0.1-4.4 kΩ with 33 mΩ(rms) resolution and higher than 97% accuracy. The proposed 25 mm² ASIC consumes only 13 μA from 2.2 V. It is a highly integrated solution offering all the functionality of acquiring multiple high quality intra-cardiac signals, requiring only a few limited numbers of external passives.
|
| [5] |
|
| [6] |
|
| [7] |
|
| [8] |
|
| [9] |
|
| [10] |
|
| [11] |
|
| [12] |
A Systematic Design Methodology for Optimization of Sigma-Delta Modulators Based on an Evolutionary Algorithm[J]. IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS, 2019, 66(9).
|
| [13] |
|
| [14] |
Analog Devices Inc.3-Channel, Low Noise,Low Power,16/24 Bit ∑-Δ ADC with On-Chip In-Amp and Reference[EB/OL].[2025-01]. https://www.analog.com/media/en/technical-documentation/data-sheets/AD7792_7793.pdf.
|
| [15] |
|
| [16] |
|
/
| 〈 |
|
〉 |